Quantitative analysis of the effect of surface treatments and designs intraocular lenses on posterior capsular opacity
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To quantitatively compare the effects of different surface treatments and designs of intraocular lens(IOL)on posterior capsule opacification(PCO)using EPCO2000.

    METHODS: A retrospective study was conducted to observe 600(971 eyes)age-related cataract patients after 1a of phacoemulsification with implantation of intraocular lens in our hospital from March to November 2016. Patients were divided into 4 groups based on the type of IOL implant: 43 eyes in ZCB00 group, 365 eyes in ZA9003 group, 340 eyes in HQ-201HEP group and 223 eyes in Human Optics group. Retroillumination slit-lamp photographs were taken after the pupil was fully dilated and the degree of PCO was scored by EPCO2000 and compared among the groups.

    RESULTS: There were 167 eyes(17.2%)with PCO involving the central 3mm of the pupil and 78 eyes(8.0%)with significant PCO or Nd:YAG laser capsulotomy. According to the hydrophobicity of IOL optical surface materials, the total score of hydrophobic group 0.000(0.000, 0.012)was obviously lower than that of hydrophilic group 0.127(0.056, 0.242)(P<0.05). The total score of heparin-free group 0.127(0.056, 0.242)was lower than that of heparin-modified group 0.175(0.067, 0.371), and the difference was statistically significant(P<0.05). The total score of single-piece group 0.000(0.000, 0.012)was obviously lower than that of three-piece group 0.120(0.041, 0.247)(P<0.05). According to the different haptic angular magnitude of IOL, the total score of the 0° haptic anglulation group 0.107(0.000, 0.212)was lower than that of the 5° haptic anglulation group 0.142(0.051, 0.298), and the difference was statistically significant(P<0.05).

    CONCLUSION: A single-piece, square-edged, hydrophobic acrylic IOL can reduce PCO formation more effectively.

    Reference
    Related
    Cited by
Get Citation

Dan-Ping Wu, Qiang Wu. Quantitative analysis of the effect of surface treatments and designs intraocular lenses on posterior capsular opacity. Guoji Yanke Zazhi( Int Eye Sci) 2019;19(11):1908-1911

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:April 19,2019
  • Revised:October 09,2019
  • Adopted:
  • Online: October 23,2019
  • Published: