Corneal biomechanical properties after femtosecond laser assisted LASIK with the corneal visualization Scheimpflug technology and ocular response analyzer
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

Supported by Xi'an Social Development Research Project Grant \〖No.SF1513(2)\〗; Shaanxi Social Development Research Project Grant(No.2016SF-274); Shaanxi Provincial Health Research Project(No.2016E002)

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To investigate the changes of corneal biomechanical properties before and after femtosecond laser assisted LASIK(FS-LASIK)using Corneal Visualisation Scheimpflug Technology(Corvis ST)and Ocular Response Analyzer(ORA), and the correlation with other myopic parameters.

    METHODS:Sixty three patients(63 eyes)who had myopic femtosecond laser assisted LASIK(FS-LASIK)were enrolled in the study. The right eye of each patient was analyzed in this study. The corneal biomechanical parameters pre-operative and 1mo post-operative was measured with the Corvis ST(Oculus, Wetzlar, Germany)and ORA(Reichert, Buffalo, New York, USA). Comparison of the biomechanical property values before and after surgery was peformed using Paired t-test or Mann-Whitney U. Pearson or Spearman correlations were used to evaluate the relationship between parameters.

    RESULTS: The postoperative 1st A-time, Vin, 2nd A length, Vout, HC time and Radius demonstrate significant decreases comparing with preoperative values(P=0.00, P=0.00, P=0.00, P=0.00, P=0.00, P=0.00 respectively). The postoperative 2nd A-time, DA and PD significantly increases(P=0.00, P=0.00, P=0.00), however, the 1st A length had no significant difference after surgery. The CH and CRF were significantly lower after FS-LASIK(P=0.00, P=0.00). A statistically significant correlation coefficient was found between preoperative central corneal thickness(CCT)with postoperative-preoperative changes of 1st A-time, 2nd A-time, DA and Radius respectively(P=0.01, P=0.04, P=0.03, P=0.01).

    CONCLUSION:There were significantly changes of corneal biomechanical properties after FS-LASIK surgery. The changes of corneal biomechanical properties after FS-LASIK can be reflected by some parameters of Corvis ST and ORA. The mainly influence of corneal biomechanical alteration was possibly correlation with corneal thickness.

    Reference
    Related
    Cited by
Get Citation

Jing Li, Sheng-Sheng Wei, Ya-Qun Wan, et al. Corneal biomechanical properties after femtosecond laser assisted LASIK with the corneal visualization Scheimpflug technology and ocular response analyzer. Guoji Yanke Zazhi( Int Eye Sci) 2017;17(2):195-199

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:November 17,2015
  • Revised:December 08,2016
  • Adopted:
  • Online: January 20,2017
  • Published: