Clinical application of the photorefraction in vision screening by the device refitted from pocket-sized digital camera
Author:
Corresponding Author:

Affiliation:

Clc Number:

R778

Fund Project:

Lianyungang Key Medical Personnel Items,China (No.ZC237)

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM:To evaluate the feasibility and reliability of the digital photorefraction device refitted from the pocket-sized digital camera in the vision screening of infants.METHODS:One hunderd infants aged from 6 months to 72 months were screened by the digital photorefraction device.The results were compared with clinical examination and retinoscopy.RESULTS:The specificities of the simple device in hyperopia,myopia,astigmatism,anisometropia,strabismus were:96.46%,95.29%,96.75%,98.97%,100%,and the sensitivities were:82.76%,93.33%,84.78%,75.00%,100%.CONCLUSION:The simple device is characterized by high sensitivity,good reliability,high specificity,simple operation,low price,easy popularization.It could be used for vision screeing of children.It is applicable to establish and popularize the family mode of vision screening.

    Reference
    Related
    Cited by
Get Citation

Bin Wang, Li Mu, Jia-Rong Zhang, et al. Clinical application of the photorefraction in vision screening by the device refitted from pocket-sized digital camera. Guoji Yanke Zazhi( Int Eye Sci) 2011;11(11):2002-2004

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:August 04,2011
  • Revised:
  • Adopted:
  • Online:
  • Published: