    Abstract:

    AIM: To compare the preoperative ocular biometry between Pentacam AXL and IOL Master 700 in cataract patients.

    METHODS:Prospective study. A total of 150 patients(150 eyes)with cataracts who were treated in our hospital from May to December 2024 were selected. The IOL Master 700 and Pentacam AXL were preoperatively used to measure axial length(AL), corneal curvature(K1, K2 and Km), anterior chamber depth(ACD), and white-to-white(WTW). The difference and consistency of the results of the two instruments were compared.

    RESULTS: There was no significant difference between the two instruments in the AL, K1, K2, Km, ACD, and WTW(all P>0.05). The Spearman correlation analysis showed that the two instruments positively correlated with the AL, K1, K2, Km, ACD and WTW of the operated eye(all P<0.001). The Bland-Altman analysis showed that for the Pentacam AXL and IOL Master 700, there were 5/150(3.3%), 7/150(4.7%), 4/150(2.7%), 5/150(3.3%), and 0 points outside the 95%LoA for the AL, K1, K2, Km, ACD, and WTW of the examined eyes, respectively, with all of these values less than 5%, indicating good consistency.

    CONCLUSION:The AL, K1, K2, Km, ACD and WTW of Pentacam AXL and IOL Master 700 in cataract patients before cataract phacoemulsification combined with IOL implantation show no significant differences, and have good correlation and consistency. The two instruments can be used interchangeably.