Comparison of central corneal thickness measurements after LASEK by three instruments
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To compare the differences and the relationships of central corneal thickness(CCT)measurements among optical coherence tomography(OCT), Tomey pachymetry and ultrasound pachymetry after laser assisted subepithelial keratomileusis(LASEK).

    METHODS:Forty-nine patients(96 eyes)with myopia after 6mo of LASEK were assigned to measure the CCT by OCT, Tomey pachymetry and ultrasound pachymetry. Paired t-test and Pearson correlation analysis was used for statistical analysis.

    RESULTS: The mean CCT measured by OCT, Tomey pachymetry and ultrasound pachymetry was(475.15±33.67μm),(447.80±34.11μm),(465.18±34.23μm), respectively. Surgery reserved mean CCT was(431.22±35.32)μm. There were statistically differences between two comparative measured results(P<0.05). Ultrasound pachymetry had a positive correlation with OCT and Tomey pachymetry(P<0.01), and the correlation coefficient was 0.949 and 0.983, respectively. The actual CCT of 6mo post-operation was thicker than surgery reserved CCT.

    CONCLUSION: The CCT measurements of three instruments were not interchangeable.

    Reference
    Related
    Cited by
Get Citation

Xiao-Juan Peng, Ye Tian, Si-Ying Liu, et al. Comparison of central corneal thickness measurements after LASEK by three instruments. Guoji Yanke Zazhi( Int Eye Sci) 2014;14(3):477-479

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:December 11,2013
  • Revised:February 14,2014
  • Adopted:
  • Online: February 27,2014
  • Published: