Clinical research of HAR-800 hand-held autorefractor on preschool children
Author:
Corresponding Author:

Affiliation:

Clc Number:

Fund Project:

Foundation of Health Science Research of the Health Bureau of Shanghai, China(No. 2008-161); Foundation of Shi-Bai-Qian Plans of Jing'an District Health Bureau of Shanghai, China(No.2010020103)

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To evaluate the accuracy of HAR-800 hand-held autorefractor on preschool children.

    METHODS:Totally 173 preschool children were included in this study. They were performed HAR-800 hand-held autorefractor examination before using atropine ointments for cycloplegic refraction. The results from HAR-800 refractor were set as experimental group, and the cycloplegic refraction were controlled group. The refractions between the two groups were compared and analyzed.

    RESULTS: The sphere was 1.59±0.61D in the experimental group, while 3.15±0.72D in the controlled group. The difference was significant(t=-82.89, P<0.001), however, there was a significant correlation(r=0.87, P<0.001, Pearsons correlation analysis). The cylinder difference was also significant(t=-6.97,P<0.001)and a significant correlation also existed between the two groups(r=0.76,P<0.001, Pearsons correlation analysis).

    CONCLUSION: The HAR-800 hand-held autorefractor can not replace cyclopegic refraction using atropine, but the results can reflect the preschool children refraction.

    Reference
    Related
    Cited by
Get Citation

Liang-Cheng Wu, Xiang-Ying Hu, Cheng-Hai Weng. Clinical research of HAR-800 hand-held autorefractor on preschool children. Guoji Yanke Zazhi( Int Eye Sci) 2012;12(12):2379-2380

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:August 08,2012
  • Revised:November 12,2012
  • Adopted:
  • Online:
  • Published: