Comparison of cornea curvature measurement by Orbscan IIz topography system and IOLMaster before refractive surgery
Author:
Corresponding Author:

Affiliation:

Clc Number:

R779.63

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM:To analyze keratometric readings obtained from Orbscan topography system and IOLMaster in myopia patients.METHODS:The corneal curvature were measured with Orbscan topography system and IOLMaster respectively in 68 cases(135 eyes) subjects aged 18-41 years old.There were 46 male(91 eyes) and 22 female(44 eyes)subjects.The results were compared.RESULTS:The average corneal curvature measured by Orbscan topography system and IOLMaster were 43.4637±1.4188 and 43.6952±1.3676D.The measurement obtained from Orbscan topography system was lower than that from IOLMaster,There is no statistical difference between the two methods.CONCLUSION:Corneal curvature abtained from IOLMaster showed no significant difference than that from Orbscan system.It is a clinically useful tool in eveluating corneal surface curvature.

    Reference
    Related
    Cited by
Get Citation

He Li, Fang-Rong Shi, Dong-Kui Zhang. Comparison of cornea curvature measurement by Orbscan IIz topography system and IOLMaster before refractive surgery. Guoji Yanke Zazhi( Int Eye Sci) 2011;11(8):1479-1480