Contrastive analysis of corneal thickness measurement difference between OCT and corneal pachymeter
Author:
Corresponding Author:

Affiliation:

Clc Number:

R779.63

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM:To investigate the clinical significance of corneal thickness measurement difference between optical coherence tomography(OCT)and corneal pachymeter.METHODS:The clinical data of corneal thickness in 102 cases 199 eyes measured by ultrasonic corneal hymeter and OCT were analyzed by statistical test.RESULTS:The thicknesses were 532.478±29.49μm and 541.47±29.00μm from OCT and ultrasonic corneal pachymeter respectively,there was statistical significance by paired t-test.It was divided into 4 groups by corneal thickness standardized on ultrasonic thickness measuring,paired t-test showed that when the corneal thickness≤500μm,500μm <corneal thickness≤550μm,550μm<corneal thickness≤600μm,there were all statistical significances between ultrasonic corneal pachymeter and OCT,when the corneal thickness>600μm,no comparison was done for little cases.CONCLUSION:Traditional ultrasonic corneal pachymeter and OCT showe some differences and better clinical conformance on measuring corneal thickness and can be used as reference before excimer laser surgery.

    Reference
    Related
    Cited by
Get Citation

Yang-Li Huang, Chi Feng, Li-Li Sun. Contrastive analysis of corneal thickness measurement difference between OCT and corneal pachymeter. Guoji Yanke Zazhi( Int Eye Sci) 2011;11(1):117-118

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:
  • Revised:
  • Adopted:
  • Online:
  • Published: