Abstract:AIM: To compare the postoperative visual outcomes and optical performance of the Eyhance ICB00 and Lentis Comfort LS-313 MF15 intraocular lenses (IOLs). METHODS: Patients who underwent phacoemulsification with bilateral IOL implantation, receiving the ICB00 IOL in one eye and the LS-313 MF15 IOL in the fellow eye, were enrolled in this retrospective study. Postoperative visual acuities, including corrected distance visual acuity (CDVA) and intermediate and near visual acuities, and monocular and binocular defocus curves were measured. Postoperative incidences of visual disturbances were also assessed. Optical bench testing evaluated the modulation transfer function (MTF) values at various defocus levels. RESULTS: The study included 52 eyes from 26 patients (16 females; 61.5%) who underwent ICB00 IOL implantation in one eye (26 eyes) and LS-313 MF15 IOL implantation in the opposite eye (26 eyes). The mean patient age was 69.2±8.4y (range, 55–84y). Both IOLs achieved excellent postoperative CDVA of 20/20 or better without significant differences between groups (88.5% vs 92.3%). Intermediate and near visual acuity were better in the LS-313 MF15 group compared with the ICB00 group, whereas the frequency of glare, starbursts, and halos was comparable between groups. The binocular defocus curve demonstrated better visual performance across defocus levels compared with both monocular curves, and the LS-313 MF15 group showed better visual acuity than the ICB00 group at negative defocus levels. Optical bench testing demonstrated that the LS-313 MF15 IOL provided higher MTF values at negative defocus levels compared with the ICB00 IOL. CONCLUSION: The LS-313 MF15 IOL demonstrates a significant advantage over the ICB00 IOL regarding intermediate and near visual acuity. The LS-313 MF15 exhibits superior optical performance on bench testing, suggesting better functional outcomes for patients requiring a broader vision range.