The binocular intraocular lens power difference in eyes with different axial lengths
Author:
Corresponding Author:

Song Chen and Xue-Feng Shi. Clinical College of Ophthalmology, Tianjin Medical University, Tianjin 300070, China. songchen1632021@163.com; shixf_tmu@163.com

Affiliation:

Clc Number:

Fund Project:

Supported by the National Natural Science Foundation of China (No.81971697; No.81501544); Shanxi Scholarship Council of China (No.2021-174); the Research Funding of Shanxi Eye Hospital (No.B201804).

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To investigate the binocular intraocular lens (IOL) power difference in eyes with short, normal, and long axial lengths (AL) using Lenstar LS 900 optical biometry. METHODS: A total of 716 (1432 eyes) participants were included. The groups were categorized into short (group A: AL<22 mm), normal (group B: 22 mm≤AL≤25 mm), and long AL groups (group C: AL>25 mm). The central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), AL, anterior corneal keratometry, white-to-white (WTW), pupil diameter (PD), as well as IOL power calculated using embedded Barrett formula were assessed. Bland-Altman plots were used to test the agreement of the binocular parameters. RESULTS: In group A, the CCT of the right eye was significantly thinner than that of the left eye (P=0.044) with a difference of -2±8 μm [95% limits of agreement (LoA), -17.8 to 13.2 μm]. For group B, the PD and IOL power in the right eye were significantly lower than those of the left eye (P=0.001, <0.001) with a difference of -0.05±0.32 mm (95%LoA, -0.68 to 0.58 mm) and -0.18±1.01 D (95%LoA, -2.2 to 1.8 D). The AL of right eye was longer than that of the left eye (P=0.002) with a difference of 0.04±0.25 mm (95%LoA, -0.45 to 0.52 mm). No significant difference was observed for all the binocular parameters in group C. The percentage of participants with binocular IOL power difference within ±0.5 D were 62% (31/50), 68.3% (339/496), and 38.8% (66/170) in groups A, B, and C, respectively. CONCLUSION: The binocular parameters related to IOL power are in good agreement, but the binocular IOL power difference of more than half of participants with long AL is more than 0.50 D.

    Reference
    Related
    Cited by
Get Citation

Ming-Hui Deng, Xiao-Gang Wang, Song Chen, et al. The binocular intraocular lens power difference in eyes with different axial lengths. Int J Ophthalmol, 2022,15(6):924-931

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:October 17,2021
  • Revised:January 28,2022
  • Adopted:
  • Online: May 31,2022
  • Published: