Comparison between two autorefractor performances in large scale vision screening in Chinese school age children
Author:
Corresponding Author:

Rui-Hua Wei and Xiao-Rong Li. Tianjin International Joint Research and Development Centre of Ophthalmology and Vision Science, Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital Tianjin, Tianjin 300384, China. rwei@tmu.edu.cn; lixiaorong@tmu.edu.cn

Affiliation:

Clc Number:

Fund Project:

Supported by Tianjin Major Science and Technology Projects (No.17ZXHLSY00070); Tianjin Clinical Key Discipline Project (No.TJLCZDXKQ012).

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To evaluate the effectiveness of Grand Seiko Ref/Keratometer WAM-5500 compared to Topcon KR800 autorefractor in detecting refractive error in large scale vision screening for Chinese school age children with the WHO criteria. METHODS: A total of 886 participants were enrolled with mean age of 9.49±1.88y from Tianjin, China. Spherical equivalent (SE) was obtained from un-cycloplegic autorefraction and cycloplegic autorefraction. Topcon KR 800 (Topcon) and Grand Seiko WAM-5500 (WAM) autorefractors were used. Bland-Altman Plot and regression were generated to compare their performance. The overall effectiveness of detecting early stage refractive error was analyzed with receiver operating characteristic (ROC) curves. RESULTS: The mean SE was -0.98±1.81 diopter (D) and the prevalence of myopia was 48.9% defined by WHO criteria according to the result of cycloplegic autorefraction. The mean SE of un-cycloplegic autorefraction with Topcon and WAM were -1.21±1.65 and -1.20±1.68 D respectively. There was a strong linear agreement between result obtained from WAM and cycloplegic autorefraction with an R2 of 0.8318. Bland-Altman plot indicated a moderate agreement of cylinder values between the two methods. The sensitivity and specificity for detecting hyperopia were 90.52% and 83.51%; for detecting myopia were 95.60% and 90.24%; for detecting astigmatism were 79.40% and 90.21%; for detecting high myopia were 98.16% and 98.91% respectively. CONCLUSION: These findings suggest that both Grand Seiko and Topcon autorefractor can be used in large-scale vision screening for detecting refractive error in Chinese population. Grand Seiko gives relatively better performance in detecting myopia, hyperopia, and high myopia for school age children.

    Reference
    Related
    Cited by
Get Citation

Di Wang, Nan Jin, Ru-Xia Pei, et al. Comparison between two autorefractor performances in large scale vision screening in Chinese school age children. Int J Ophthalmol, 2020,13(10):1660-1666

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:April 13,2020
  • Revised:May 18,2020
  • Adopted:
  • Online: August 25,2020
  • Published: