Effect of femtosecond and microkeratome flaps creation on the cornea biomechanics during laser in situ keratomileusis: one year follow-up
Author:
Corresponding Author:

Yue-Hua Zhou; Xiao-Yan Peng. Ophthalmic Center, BeijingTongren, Hospital, No. 1 Dongjiaomin Ln, Dongcheng District, Beijing100730, China. yh0220@yahoo.com

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To compare the corneal biomechanical outcomes at one year after laser in situ keratomileusis (LASIK) with the flaps created by Ziemer and Moria M2 microkeratome with 110 head and -20 blade. METHODS: Totally 100 eyes of 50 consecutive patients were enrolled in this prospective study and divided into two groups for corneal flaps created by ZiemerFemto LDV and Moria M2 microkeratome with 110 head and -20 blade. Corneal biomechanical properties including cornea resistance factor (CRF) and cornea hysteresis (CH) were measured before and 1, 3, 6, 12mo after surgery by ocular response analyzer. Central cornea thickness and corneal flap thickness were measured by optical coherence tomography. RESULTS: The ablation depth (P=0.693), residual corneal thickness (P=0.453), and postoperative corneal curvature (P=0.264) were not significant different between Ziemer group and Moria 110-20 group after surgery. The residual stromal bed thickness, corneal flap thickness, CH and CRF at 12mo after surgery were significant different between Ziemer group and Moria 110-20 group (P<0.01);Ziemer group gained better corneal biomechanical results. The CRF and CH increased gradually from 1 to 12mo after surgery in Ziemer group, increased from 1 to 6mo but decreased from 6 to 12mo in Moria 110-20 group. Both CRF and CH at one year after surgery increased with the increasing of residual cornea thickness; pre-LASIK CRF, CRF also increased with residual stromal bed thickness, while CH decreased with the increasing of pre-LASIK intraocular pressure and cornea flap thickness (P<0.01). CONCLUSION: In one year follow-up, femtosecond laser can provide better cornea flaps with stable cornea biomechanics than mechanical microkeratome.

    Reference
    Related
    Cited by
Get Citation

Qian Sun, Zheng-Zheng Deng, Yue-Hua Zhou, et al. Effect of femtosecond and microkeratome flaps creation on the cornea biomechanics during laser in situ keratomileusis: one year follow-up. Int J Ophthalmol, 2016,9(10):1409-1414

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
Publication History
  • Received:February 07,2015
  • Revised:July 27,2015
  • Adopted:
  • Online: October 13,2016
  • Published: