Agreement of corneal high-order aberrations measured by three wavefront aberrometers in myopic adults
Author:
Affiliation:

Fund Project:

Guangxi Science and Technology Project(No.GuiKeAD19245193)

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
  • |
  • 文章评论
    摘要:

    目的:评估Topcon KR-1W、i.Profiler、OPD-Scan Ⅲ三种波前像差仪测量成年近视患者角膜高阶像差的一致性。

    方法:前瞻性临床研究。收集2022-06/08在广西壮族自治区人民医院视光中心就诊的成年近视患者92例92眼,分别采用Topcon KR-1W、i.Profiler、OPD-Scan Ⅲ测量4、6mm瞳孔直径下角膜三、四阶像差,评估三种波前像差仪的差异性和一致性。

    结果:三种像差仪6mm瞳孔直径下测量值均高于4mm瞳孔直径下测量值。4mm瞳孔直径下三种像差仪Z-44、Z-24测量值均无差异(P>0.05),其余测量值均有差异(P<0.05); 6mm瞳孔直径下三种像差仪测量值均有差异(P<0.05)。4mm瞳孔直径下三阶像差测量值、6mm瞳孔直径下三、四阶像差(除Z-24)测量值的95%LoA>0.1μm,一致性相关系数Pc<0.90,显示一致性较差。三种像差仪4、6mm瞳孔直径下角膜高阶像差测量值相关性差异较大(r4mm=0.215~0.805,P4mm<0.05; r6mm=0.561~0.916,P6mm<0.001)。

    结论:Topcon KR-1W、i.Profiler与OPD-Scan Ⅲ测量4、6mm瞳孔直径下角膜三、四阶像差时测量结果存在差异性,一致性较差,临床应用中三种像差仪不可相互替代。

    Abstract:

    AIM:To evaluate the agreement of corneal high-order aberrations from Topcon KR-1W, i.Profiler and OPD-Scan Ⅲ wavefront aberrometers in myopic adults.

    METHODS:A prospective clinical study. A total of 92 adult patients(92 eyes)with myopia in the department of optometry, the People's Hospital of Guangxi Zhuang Autonomous Region from June to August 2022 were enrolled. The third-order and fourth-order corneal aberrations at the pupil diameter of 4 and 6mm were measured by Topcon KR-1W, i.Profiler, and OPD-Scan Ⅲ, respectively. The difference and agreement of the three aberrometers were evaluated.

    RESULTS: The measurements at 6mm pupil diameter were all greater than those at 4mm pupil diameter. Although there were no statistical differences in the measurements of Z-44、Z-24 by the three aberrometers at 4 pupil diameter(P>0.05), there were statistical differences in other measurements(P<0.05). The aberration results measured by the three aberrometers were statistically different at the 6mm pupil diameter(P<0.05). The 95% limit of agreement(95%LoA)of the measurements of higher-order aberration, including the third-order aberrations at 4mm pupil diameter and the third-order and fourth-order aberrations at 6mm pupil diameter(except for the Z-24)were greater than 0.1μm. The concordance correlation coefficient(Pc)was lower than 0.90, indicating a poor consistency. The correlation coefficients of corneal higher-order aberrations were significantly different among the three aberrometers at 4 and 6mm pupil diameter(r4mm=0.215~0.805, P4mm<0.05; r6mm=0.561~0.916, P6mm<0.001).

    CONCLUSION:There were significant differences in the measurements of the third- and fourth-order corneal aberrations at 4 and 6mm pupil diameter among Topcon KR-1W, i.Profiler, and OPD-Scan Ⅲ, and the agreements were poor, so they are not interchangeably in clinical applications.