[关键词]
[摘要]
目的:运用新型光学生物测量仪IOL Master 700测量白内障超声乳化手术前后眼部生物学参数的变化,并探讨人工晶状体(IOL)屈光度数计算公式的选择。
方法:前瞻性研究。收集2021-01/06在苏州大学附属第一医院就诊的白内障患者52例57眼。术前和术后3mo使用IOL Master 700完成眼轴长度(AL)、前房深度(ACD)、角膜曲率(Km)的测量并分析。对不同IOL公式计算时预留的目标屈光值与术后3mo全自动验光仪实际屈光值结果进行比较并分析。
结果:手术前后测量的AL平均值分别为24.20±1.86、24.09±1.86mm,术后AL缩短了0.11mm; ACD值分别为3.08±0.44、4.55±0.36mm(P<0.001),术后ACD加深1.49mm; Km值分别为44.14±1.86、44.14±1.82D(P>0.05)。术前选用Barrett Universal Ⅱ公式所测结果的屈光误差最小,其次是Holladay Ⅱ及SRK/T公式,Holladay Ⅰ公式所测结果的误差最大(P<0.05)。
结论:白内障术后AL缩短以及ACD加深,度数测算时可考虑增加0.1mm的校正因子。IOL屈光度数计算公式中Barrett Universal Ⅱ公式预测性最佳,其次是Holladay Ⅱ及SRK/T公式。
[Key word]
[Abstract]
AIM: To measure the changes of ocular biological parameters before and after phacoemulsification, and compared the choice of intraocular lens(IOL)power calculation formulas based on the new optical biometric instrument IOL Master 700.
METHODS: A prospective study. Clinical data were collected from 52 patients(57 eyes)with cataract at the First Affiliated Hospital of Soochow University from January to June 2021. The axial length(AL), anterior chamber depth(ACD)and corneal curvature(Km)were measured and analyzed before and 3mo after phacoemulsification by IOL Master 700. The target refractive value reserved in the calculation of different IOL formulas and the actual refractive value of the automatic refractor 3mo after phacoemulsification were compared and statistically analyzed.
RESULTS: The average values of AL measured before and after phacoemulsification were 24.20±1.86, 24.09±1.86mm, the postoperative AL shortened by 0.11mm, and the ACD values were 3.08±0.44, 4.55±0.36mm(P<0.001), ACD deepened by 1.49mm after phacoemulsification. The Km values were 44.14±1.86, 44.14±1.82D(P>0.05). The refractive error of the results measured by the Barrett Universal Ⅱ formula was the smallest before operation, followed by Holladay Ⅱ and the SRK/T formula, the Holladay Ⅰ formula had the largest error and the difference was statistically significant(P<0.05).
CONCLUSION: The AL was shortened and the ACD was deepened after phacoemulsification. A correction factor of 0.1mm is suggested to add when calculating the degree. The Barrett Universal Ⅱ formula has the best predictability in the IOL power calculation formulas, follow by Holladay Ⅱ and SRK/T formula.
[中图分类号]
[基金项目]
国家自然科学基金面上项目(No.81970830)